Home. 분석서비스. Analytes & Panels
Configurable-Human Kidney Biomarker Panel
페이지 정보
본문
Target Analytes
Analyte | Standard Curve(pg/mL) | Serum | Plasma | CellCulture | Sensitivity(pg/mL) |
Clusterin | 664 - 483,800 | 1:4,000 | 1:4,000 | N/R | 64.3 |
CXCL10/IP-10/CRG-2 | 4.27 - 3,110 | 1:10 | 1:10 | N/R | 0.693 |
Cystatin C | 67.9 - 16,500 | 1:4,000 | 1:4,000 | N/R | 57.1 |
Lipocalin-2/NGAL | 49.5 - 36,100 | 1:10 | 1:10 | N/R | 63.1 |
Osteopontin/OPN | 405 - 295,500 | 1:10 | 1:10 | N/R | 247 |
TFF3 | 15.1 - 11,000 | 1:10 | 1:10 | N/R | 11.5 |
* Clusterin requires at least a 1:4000 dilution when assaying serum and plasma samples.
* Cystatin C requires at least a 1:4000 dilution when assaying serum and plasma samples.
* Urine samples require a 10-fold dilution. A suggested 10-fold dilution is 20uL of samples + 180uL of Calibrator Diluent RD6-62 (diluted 1:5).
특징
- 6가지 human kidney biomarker중 원하는 타겟을 선정할 수 있는 multiplex performance kit
- 소량의 샘플 이용 (<50ul)
- 재현성 있는 결과를 위해 mass-calibrated standard 제공
- 손쉬운 wash step을 위한 magnetic format
- R&D Systems의 Quantikine ELISA Kit와 비슷한 수준의 validation testing을 거쳐 생산됨
Kit components
- Premixed cocktail of antibody-coated Magnetic beads
- Premixed cocktail of biotinylated detection antibody
- Standard Cocktail(s)
- Bead Diluent
- Biotin Antibody Diluent
- Calibrator Diluent
- Wash Buffer
- Streptavidin-PE
- One flat-bottom 96-well Microplate
- Foil Plate Sealers (4)
- Mixing Bottle
- Standard Value Card (A standard curve must be generated each time an assay is run, utilizing values from the Standard Value Card included in the Kit.)
-
- 이전글
- Configurable-Human MMP Panel
- 24.05.09
-
- 다음글
- Configurable-Human Immuno-Oncology Panel
- 24.05.09
댓글목록
등록된 댓글이 없습니다.
단백질전기영동 & Precast Gel, 사이토카인 ELISA kit & 분석서비스, Direct PCR
COMPANY INFO
회사명: LABIS KOMA
통신판매업 신고번호:
주소 : 서울시 영등포구 양평로21길 26, 아이에스비즈타워 1907
CONTACT DETAILS
T : 02-6245-5500
F : 02-6248-8800
E : support@labiskoma.co.kr
2024 © LABIS KOMA. All Rights Reserved.